© Siemens Laser Analytics Jan 2001 CD 3002 Flange installation manual Document number: CID 3002-1107 Rev. 1C

LDS 3000 Flange installation

Content

Content 1. Installation of flanges for CD 3002

1

1.1 General 1.2 Installation 1.3 Measurement path length 1.4 Path length simulations

1 1 3 3

I

Content

II

LDS 3000 Flange installation

LDS 3000 Flange installation

Installation of flanges for CD 3002

1. Installation of flanges for CD 3002 1.1 General Before CD 3002 can be used a pair of flanges must be installed at the measurement site. The flanges must be installed opposite each other at an angle error of not more than ±2° relative the mutual axis of symmetry.

1.2 Installation The flanges should be welded to the measurement volume according to the figure below. In some cases (when the walls of the measurement volume is not rigid enough) an external framework that holds the flanges is necessary. The stability of the flange position must be such that the extended axis of symmetry for one flange does not move more than ±10 mm at the plane of the other flange. The stability requirement consequently becomes more critical for longer measurement paths.

±1 cm

Max dynamic movement