tected mode should be done by trained service technicians. Malfunction due to faulty settings are not covered by the warranty. 3.1 Temperature setting.
3. Settings 3.1 Temperature setting 3.2 Over temperature protection 3.3 Tuning 3.4 Parameter list for Omron E5CN Temperature controller
1 1
3 3 3 3 4
5 5 5 6 7
I
Content
II
LDS 3000 Operators manual
LDS 3000 Operators manual
AH 3000
1. AH 3000 1.1 General The optical surfaces on the sensors are in many applications exposed to very nasty environments and will get stained very quickly if no measures are taken. The most common method to protect the optical surfaces is by flushing purging gas in front of them. In some applications, like in waste combustion, where a large number of reactive gas components are present the purging gas must be heated to prevent condensation of salts on the lens surface. To do this the air heater AH 3000 can be used. Isolated hot air tube
Purge air input Low air temperature indicator Power indicator
High air temperature indicator
Mains switch Mounting bracket
AH 3000.
The mains switch turns the heater on and off. The power indicator (white) indicates that the heater is operating. The low air temperature indicator (yellow) indicates that the temperature of outgoing air is lower than its set value. The high air temperature indicator (red) indicates that the temperature of the outgoing air is higher than its set value. In the start-up phase both high and low temperature indications are normal, until the temperature is properly regulated to the set value
1
AH 3000
2
LDS 3000 Operators manual
LDS 3000 Operators manual
Installation
2. Installation 2.1 Mechanical Fix the heater to wall or brackets with M8 bolts or screws of equivalent strength. Note that the casing can become hot (up to 80 °C). Cabinet size is 400x300x160 mm (HxWxD) on the right side 100 mm has to be added for the air connections and at the bottom a further 70 mm is needed for the electrical connection. The bolt position should be as below:
268 mm
8 mm
70 mm
435 mm
200 mm
2.2 Pressurized air Connect tubes for incoming and outgoing pressurised air. Both intake and output are ¼” thread. The pressurised air MUST be OIL FREE with a pressure of 4-8 bars and with a maximum flow rate of 500 ltr/min.
2.3 Electrical There are two types of heater, one for 110V/60Hz and one 230V/50Hz. Make sure the heaters are connected to the right voltage. Current rating: 9A/230V, 19A/110V
3
Installation
LDS 3000 Operators manual
Connected power: 2kW The heater is supplied with a power plug complying to the standard in the destination country. Note: If the power plug is not used note that in most countries the electrical installation has to be done by a certified electrician.
2.4 Insulation Insulate tubing to avoid any heat losses, take extra care with metal connectors since the heat loss is greatest from these.
4
LDS 3000 Operators manual
Settings
3. Settings Introduction Settings can be done in three modes of the temperature controller - Operation level, Adjustment level and Protected level. All settings in protected mode should be done by trained service technicians. Malfunction due to faulty settings are not covered by the warranty.
3.1 Temperature setting This setting is done under Operation level. In the normal operation mode the keys 3 and 4 are used to set the temperature. 3 to decrease the set value and 4 to increase it.
3.2 Over temperature protection This setting is done under Protected level. The over temperature protection is a floating value that depends on the set temperature. Normally this value is set to -2 which means 2 °C above the set temperature. If there is a low airflow and/or low temperature set it might be necessary to increase the span for the over temp protection and in some cases even the low temperature indication. To do so follow these steps:
Enter Protected level: 1. 2. 3. 4. 5. 6.
Keep keys 1 and 2 pressed for at least 1 second. Toggle through the different parameters with key 2. Change öAPt to 0 Change iCPt to 0 Check that utPt is set to OFF. Return by keeping keys 1 and 2 pressed for at least 1 second.
Set the temperature protection: 1. Toggle through the different parameters with key 2. 2. r-5 should always be set to run. Setting is done with keys 3 and 4. 3. AL-1 is the low temperature limit normally set to 10 °C. This means that the low temperature indicator will light if the temperature goes below set value by 10 °C. Note the polarity. Setting is done with keys 3 and 4. 4. AL-2 is the over temperature protection normally set to -2. This means that the heating is shut off if the temperature exceeds the set value with more than 2 °C. Operation will proceed when the temperature has dropped back to the set value. Note the reversed polarity. Setting is done with keys 3 and 4.
Exit Protected level: 1. 2. 3. 4.
Keep keys 1 and 2 pressed for at least 1 second. Toggle through the different parameters with key 2. Change öAPt to 2 Change iCPt to 2
5
LDS 3000 Operators manual
Settings
5. Check that utPt is set to OFF. 6. Return by keeping keys 1 and 2 pressed for at least 1 second.
3.3 Tuning This setting is done under Protected level. An alternative to changing the upper and lower limit when there is problem with regulating the temperature to the set value is performing an “Auto-tune”. Auto-tune will run several heating cycles and then automatically set new PID-parameters. Before running auto-tune the pressurised air must be connected and the temperature must be set to the wanted value - AH 3000 must be in normal operating condition.
Enter Protected level: 1. 2. 3. 4. 5. 6.
Keep keys 1 and 2 pressed for at least 1 second. Toggle through the different parameters with key 2. Change öAPt to 0 Change iCPt to 0 Check that utPt is set to OFF. Return by keeping keys 1 and 2 pressed for at least 1 second.
Then start auto-tune by: 1. 2. 3. 4.
Press key 1. Set At to ON. The temperature will flash during execution of auto-tune. When auto-tune is done the temperature stops flashing, it will automatically be set to OFF and the new PID-parameters will be stored. 5. Exit the adjustment level by pressing key 1.
Exit Protected level: 1. 2. 3. 4. 5. 6.
6
Keep keys 1 and 2 pressed for at least 1 second. Toggle through the different parameters with key 2. Change öAPt to 2 Change iCPt to 2 Check that utPt is set to OFF. Return by keeping keys 1 and 2 pressed for at least 1 second.
LDS 3000 Operators manual
Settings
3.4 Parameter list for Omron E5CN Temperature controller Metric
Slide the sensor card into its holder on the calibration cell module and connect the power cable from the DC/DC card. Insert the detector mod- ule into the hole ...
number of methods available to keep the lenses in good shape. They will .... Siemens Laser Analytics AB. Setup for purging the lens with steam. Air buffer.
4.4 Adjustment of AGC-card. 19. 4.5 Save ... just the angle of the flange tube until the light spot is centred on the ..... On systems with sensor cards older than ver.
The LDS 3000 is a system for on-line gas analysis providing continuous presentation of .... Thus the sensor optics can very easily be cleaned if needed. ... flue gas or contamination of the optical components, is compensated for automatically to ...
Protective gloves should be ised when it is handled. Halogen ... Twist-mount the alignement lamp instead of the cross hair in the transmitter. Transmitter side.
Setting-up a MSAccess 2000 database for a MSWindows operating system ...... The report answers the question, 'if the period in question ended now, what.
components and safety features to precisely control combustion and minimize the ..... Be sure to install an external temperature sensor if air heaters are operated in ...... Although simple maintenance procedures can be performed by the owner, ...
Diagnoisis and repair of mafunctioning, non-functioning or damaged Webasto ..... Webasto recommended maintenance procedures per Webasto Owners Manual. ..... air and heated air ducting, and most importantly, the exhaust line maximum ...
26 mai 2015 - Close the window. 4. Increase the temperature setting. 5. Select the higher power setting. 2. The set temperature is lower than the current room.
For filling and bleeding of hydraulic brakes. ¤ Brand new in aviation ... The innovation is worldwide filed for patent. Sole distributor is Tost Flugzeuggerätebau ...
power still applied. When opening covers or ... Overview. Description. This manual provides instructions for installation of the Maxum edition ..... If a forklift or dolly is used, make sure the Maxum II is strapped to it se- curely. Add padding ..
The HydroRanger 200 is a single-point, six relay device that offers both level and volume ... è Wall Mount: suitable for outdoors (Type / Nema 4x, IP65 Enclosure).
24, ost SA locking cevice. 27, Air evoCuction CDrtoiner ... Note the side clistance oetween or evacuation CDntainer and probe, ca, 500mm. Datur 2, 2,96. Bedro ...
be construed as a commitment, representation, warranty, or guarantee of any method, ...... Once your PC is set up, as described in Configuring Your PC, you are.
Description. In the methanator CO and CO2 are ... Methanator Operation. Ni. CO + 3 H2. â. CH4 + H2O. Ni ... Operational Note. File: Filament.DOC. Removal ...
This device/system may only be set up and operated in conjunction with this manual. ... Under Weighing Technology, select Continuous Weighing Systems and then go to the manual ...... The distance calculation is dependent upon the ambient sound. .....
>Many many thanks again, Stealth Pilot. >. . . I don't even feel I need the Cessna manual as your description was so. >clear and concise. I feel we'll get it done ...
Neopren. Rubber bushing. EPDM 50. Packaging. Bio-degradable cardboard. 3K. âÎ. 2K. âÎ ... However, the data in this manual are reviewed regularly and any.