Tobramycin by LCMS - Microsolv

Uracil (m/z 113), Tobramycin (m/z 468). 1 mg/L of each dissolved in DI water. Detection: Mass Spec: Atmospheric Pressure. Chemical Ionization in positive mode ...
80KB taille 2 téléchargements 214 vues
HPLC

APPLICATION NOTES

TM

UDC Cholesterol

TM

AminoGlycoside: Tobramycin by LCMS OH

NH2

H O

HO

O

NH2

O

H H2N

NH2

H2N

Method Conditions

OH

O

OH OH

Column: Catalog No.: Dimensions: Mobile phase:

Cogent UDC Cholesterol™, 4µm, 100Å. 69069-7.5P 4.6 x 75 mm 60:40 acetonitrile/DI water + 0.5% formic acid Temperature: Ambient pH: 2.90 Flow rate: 0.5 mL/minute Injection Volume: 1 µL Samples: Uracil (m/z 113), Tobramycin (m/z 468) 1 mg/L of each dissolved in DI water Detection: Mass Spec: Atmospheric Pressure Chemical Ionization in positive mode: APCI+ Single Ion Monitoring

Discussion Tobramycin is a drug used for treatment in patients with cystic fibrosis and other diseases caused by gram-negative bacteria. It has a very narrow therapeutic range, so monitoring of the drug in patients’serum is crucial to avoid devastating side effects. A simple, isocratic method for analyzing Tobramycin is presented: Note: Aminoglycoside drugs are difficult to retain on conventional HPLC columns because of their highly polar characteristics. Ion pair reagents have to be used in order to induce retention, but they interfere with MS detection. UV detection requires derivatization of the drug. No ion pair agents were used in the above.

Cat. No. 69069-7.5P

NO DERIVATIZATION AND NO ION PAIR REAGENTS ARE USED IN THE METHOD. Symmetrical peak shape and reasonable retention make this method an excellent choice for an assay of this drug. For more information visit www.MTC-USA.com

Description Cogent UDC-Cholesterol™ HPLC Column, 100Å 4µm, 4.6mm x 75mm, Standard End Fittings. APP B05-2

MTC • MicroSolv Technology Corporation 1 Industrial Way West, Bldg E, Eatontown, NJ 07724 USA Ph. 1-732-578-1777 • Fx. 1-732-578-9777 www.mtc-usa.com Email: [email protected]