LINEARITY : < +/- 0.5 % (F.S.O) OF THE PROBE RANGE ... DYNAMIC RESOLUTION AT MAXIMUM PROBE DISTANCE : 0.15 µm with 1Hz ... VIBRATION.
131KB taille 2 téléchargements 268 vues
STATIC AND DYNAMIC NO CONTACT DISTANCE MEASUREMENT CAP 100

& CAP 100 / 2 / 4

CAPACITIVE CONDITIONING UNIT

INDUSTRIAL CASING (1 / 2 / 4 channels)

• MEASURING RANGE : FROM 0→ 0.5 mm to 0 → 12 mm according to the diameter of the sensor electrode used • LINEARITY : < +/- 0.5 % (F.S.O) OF THE PROBE RANGE • TEMPERATURE DRIFT (/ °C) < 0.005 % (F.S.O) OF PROBE RANGE • FREQUENCY RANGE (-3dB) : 0 to 20 kHz typical • DYNAMIC RESOLUTION AT MAXIMUM PROBE DISTANCE : 0.15 µm with 1Hz analysis bandwidth and 15 µm with 10 kHz bandwidth (no averaging) • MEASUREMENT OUTPUT : 0 / +10V LINEAR VARIATION Vs DISTANCE OR CAPACITANCE (selection when ordering) • OFFSET ADJUSTEMENT: FROM 0 to -10V (on option) • POWER :5 watt ; 230 Vac or 12 to 24 V DC (selection when ordering) • PACKAGING : Industrial casing IP 64 (dimensions : 180*135*70 for 1 channel unit) • COMPATIBLE WITH CAPAAB or OTHER BRAND CAPACITIVE TRIAXIAL PROBES (Temperature range: from -270°C to +800°C)

Example of 4 channels industrial casing with LED display of static distance (on option)

______________________________________________________________________________________ CAPAAB 22/32 sentier des Torques BAT 8 92290 Chatenay Malabry www.capaab.fr France

CAPACITIVE STANDARD PROBE SIZE REFERENCE DISTANCE RANGE (mm) EXTERNAL DIAMET. (mm) LENGHT (mm)

CC1 1

CC2 2

CC5 5

CC8 8

CC12 12

4

6

15

24

36

23

25

28

20 20 Temperature range : -55 °C to +125 °C Probe cable standard length : 2 meters

TYPICAL APPLICATIONS SINGLE CHANNEL MEASUREMENT

DISPLACEMENT

EXCENTRICITY

VIBRATION

OIL FILM THICKNESS

CASE TO BLADE TIP CLEARANCE

BADE TIP CLEARANCE + PRECESSION

SEVERAL CHANNEL MEASUREMENTS

CONCENTRICITY

CONDUCT. LIQUID HEIGHT (I.E WATER)

ALIGNMENT

POSITIONING

DIAMETER MEASUREMENT

SHAFT ORBITAL MOVEMENT

TIP BLADES ORBITAL MOVEMENT

THICKNESS MEASUREMENT

CAPAAB 22/32 sentier des Torques BAT 8 92290 Chatenay Malabry France www.capaab.fr Phone + 33 1 71 22 73 72